Measurement of the performance of sound level meters using IEC 61672-3:2006

Auteurs-es

  • Peter Hanes Institute for National Measurement Standards, National Research Council Canada, 1200 Montreal Rd., Ottawa, Ont. K1A 0R6, Canada

Mots-clés :

Acoustic signal processing, Calibration, Mathematical models, Calibrated sound calibrators, Electrical signals, Single frequency

Résumé

IEC 61672-3:2006 provides the first internationally agreed methods for measuring the performance of individual sound level meters (SLM). Pattern approval tests are specified in IEC 61672-2, but are in practice only performed in countries where legal requirements exist for testing a model of SLM before it may be sold. A calibrated sound calibrator is applied in order to determine the overall sensitivity at a reference level at a single frequency. The indicated sound level with no input signal is noted; this corresponds to the level of self-generated noise in the SLM. The indicated peak C sound level in response to single cycle and half cycle sinusoidal electrical signals is measured and compared with the theoretical ideal response. IEC 61672-3 should be recommended for use in Canada with appropriate guidance on its use with respect to older SLMs.

Fichiers supplémentaires

Publié-e

2007-09-01

Comment citer

1.
Hanes P. Measurement of the performance of sound level meters using IEC 61672-3:2006. Canadian Acoustics [Internet]. 1 sept. 2007 [cité 15 mai 2026];35(3):144-5. Disponible à: https://jcaa.caa-aca.ca/index.php/jcaa/article/view/1942